BWSK Calibration and adjust

BWSK Calibration and adjust



To calibrate and adjust the temperature reading of a BWSK.

From the main menu, select  “Devices” (1). Then, from the emerging menu select “Device List” (2).
Device listFind in the device list the BWSK sensor to work with, and select the settings button (1).

Device settingsExpand the “Factor”(1) parameter from the settings list and copy the current value (2).

Copy factor
Expand the “Offset”(1) parameter from the settings list and copy the current value (2).

Copy offset

Paste the current factor and current offset in the corresponding fields in the following form.
Current Factor:
Current Offset:
Set the test temperature to the lowest value of the desired range and wait until the readings from the BWSK and the reference thermometer stabilize.
Enter in the following field the reading from the BWSK.
BWSK low temperature:
Enter in the next field the temperature of the reference thermometer.
Reference low temperature:
Set the test temperature to the highest value of the desired range and wait until the readings from the BWSK and the reference thermometer stabilize.
Enter in the following field the reading from the BWSK.
BWSK high temperature:
Enter in the next field the temperature of the reference thermometer.
Reference high temperature:


From the device settings, expand the “Factor”(1) setting and copy the new calculated factor into the “Set Value”(2) field. Save the new factor by selecting “Submit”(3)

Save factor
From the device settings, expand the “Offset”(1) setting and copy the new calculated offset into the “Set Value”(2) field. Finally, save the new offset by selecting “Submit”(3).

Save offset